Frequently Asked Questions

 

Add new question

1 | 2 | 3 | 4 | 5 | 6 | next »

FAQs Overview

  • What is the space requirement for MTP instruments?

    MTP instrument including travel arrangement width is 180 mm and base plate dimension is 245 mm under the strip (rarely 190 mm). Beyond or ahead of the strip a space must be provided for operating movement of the instrument in accordance to submitted design. Drawing showing the outside dimensions, instrument's travel arrangement and evaluating unit will be prepared and submitted after an agreement on the type of measuring unit required is reached.

  • What is included in the MTP instrument delivery?

    Delivery includes the sensor head with positioning system, evaluation unit with digital and analogue display, and pneumatic distributor with air beneficiation, cables & pipes for interconnections. The client provides the base plate under the positioning system in accordance with the supplied documentation.

  • When is it advantageous to use cheaper MTUP?

    If the strip width is below 50 mm and flat. MTUP instrument (Continuous Narrow Strip Thickness Gauge) is not equipped by "swinging head" as MTP gauge (3 axis discretion) but only by height adjustment (1 axis discretion).

  • Is it possible to measure sheet thickness?

    No. The instruments are used for a continuous measurement of strip thickness.

  • Is it possible to use parts from old original instruments from other manufacturers?

    No, due to different designs it is not possible to use any part of the old or obsolete devices from other manufacturers.

  • How the instruments are priced and what are the purchase requirements?

    As a minimum it is necessary to state the manual and/or automatic option, the required distance of measurement from the strip edge (130; 210; 310 mm), any other special requirements for serial data output and equipment of outputs and accessories (pen recorder, collection of data with report printing tc). Further more, minimum and maximum of strip width and thickness, type of material, speed of the strip and positioning requirement.

  • How often is the instrument calibrated, and who does it?

    Once a year, provided by UVB TECHNIK appointed personnel or trained local agent technicians or customer´s metrology representative.

  • What is necessary to change regularly on the instruments?

    Air filters must be cleaned or changed regularly.

  • Are any special measuring or tools requirements are necessary for common reparations and adjustments?

    For normal - common - adjustments and reparations except digital voltmeter and etalons of thickness for calibration purposes (etalons are part of delivery) no other special instruments are needed.

  • Who provides the service?

    Training of operators and maintenance is a part of installation procedure; mostly operating personnel does the local maintenance, which is not complicated. E-mail or faxed requirements are detailed explained and if necessary our or local agent technicians are available.